MEDISENSE OPTIUM
Report
- Report Number
- 2954323-2011-00741
- Event Type
- Malfunction
- Date Received
- January 19, 2011
- Date of Event
- December 18, 2010
- Report Date
- January 19, 2011
- Product Code
- NBW
- Removal / Correction Number
- ADC FA1197-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30C DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE C, D OR E ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. THE PRIMARY CAUSE HAS BEEN IDENTIFIED TO BE BATCH TO BATCH GLUE VARIABILITY. THE FDA HAS BEEN INFORMED OF THE FIELD ACTION PER 21CFR806 (FIELD ACTION REFERENCE NUMBER 2954323-12/22/10-001-R). ALL AFFECTED CONSIGNEES WERE NOTIFIED BY LETTER BEGINNING DECEMBER 22, 2010.
IN USING AN AFFECTED TEST STRIP RELATED TO AN ON-GOING FIELD ACTION FOR ABBOTT'S PRECISION FAMILY TEST STRIPS, THE CUSTOMER REPORTED EITHER LONGER BLOOD FILL TIME OR READINGS ISSUES. THE CUSTOMER REPORTED GETTING LOWER THAN FEELS READINGS, EXPERIENCING SYMPTOMS OF HYPOGLYCEMIA WITH SUBSEQUENT LOSS OF CONSCIOUSNESS AND SEIZURE. THE CUSTOMER REPORTEDLY WAS TREATED WITH GLUCOSE INTRAVENOUS INFUSION BY PARAMEDICS AND DRANK MILK TO COUNTERACT THE EVENT. THE EVENTS OCCURRED ON (B)(6) 2010 AT 4:45PM AND (B)(6) 2010 7:00PM. THERE IS NO INDICATION ADC DEVICE CONTRIBUTED TO THE MEDICAL EVENT, AS THE REPORTED READINGS OF 45 MG/DL WERE CONSISTENT WITH THE TREATMENT PROVIDED. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS MEDICAL EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDISENSE OPTIUM | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 45841 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |