FDA Adverse Event Malfunction Summary report: N

R.O. WATER TREATMENT

MDR report key: 1964302 · Received January 19, 2011

Report

Report Number
1423500-2011-00785
Event Type
Malfunction
Date Received
January 19, 2011
Date of Event
December 27, 2010
Report Date
December 27, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FIP
PMA / PMN Number
K822385
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT SAMPLE HAS BEEN REQUESTED BUT HAS NOT YET BEEN RECEIVED BY BAXTER FOR EVALUATION. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL RO DEVICE WAS EVALUATED AND THE REPORTED CONDITION: OF A BURNING SMELL WAS CONFIRMED. THE ROOT CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE THE MOTOR. THE MOTOR AND PUMP HEAD WERE REPLACED. THE DEVICE WAS TESTED AS PER BAXTER OPERATING PROCEDURES AND PROTOCOLS AND PASSED ALL TESTING. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER (B)(4) TO REPORT THAT THE DEVICE HAD A BURNING SMELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 R.O. WATER TREATMENT SUBSYSTEM, WATER PURIFICATION FIP BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1