FDA Adverse Event
Malfunction
Summary report: N
R.O. WATER TREATMENT
MDR report key: 1964302
·
Received January 19, 2011
Report
- Report Number
- 1423500-2011-00785
- Event Type
- Malfunction
- Date Received
- January 19, 2011
- Date of Event
- December 27, 2010
- Report Date
- December 27, 2010
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FIP
- PMA / PMN Number
- K822385
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4)
Additional Manufacturer Narrative · 1
(B)(4). THE PRODUCT SAMPLE HAS BEEN REQUESTED BUT HAS NOT YET BEEN RECEIVED BY BAXTER FOR EVALUATION. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative · 1
(B)(4). THE ACTUAL RO DEVICE WAS EVALUATED AND THE REPORTED CONDITION: OF A BURNING SMELL WAS CONFIRMED. THE ROOT CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE THE MOTOR. THE MOTOR AND PUMP HEAD WERE REPLACED. THE DEVICE WAS TESTED AS PER BAXTER OPERATING PROCEDURES AND PROTOCOLS AND PASSED ALL TESTING. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BAXTER (B)(4) TO REPORT THAT THE DEVICE HAD A BURNING SMELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | R.O. WATER TREATMENT | SUBSYSTEM, WATER PURIFICATION | FIP | BAXTER HEALTHCARE - LARGO |