FDA Adverse Event Injury Summary report: N

XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 1964298 · Received January 19, 2011

Report

Report Number
2024168-2011-00380
Event Type
Injury
Date Received
January 19, 2011
Date of Event
April 28, 2010
Report Date
December 25, 2010
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P070015
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE 3.5 X 15 MM XIENCE V (1009542-15, 7110641) IS BEING REPORTED UNDER A SEPARATE MEDWATCH REPORT NUMBER. ABOVE RATED BURST PRESSURE, INDICATION FOR USE. THERE WAS NO REPORTED PRODUCT DEFICIENCY. ANGINA, MYOCARDIAL INFARCTION, AND RESTENOSIS ARE KNOWN ADVERSE EVENTS AS LISTED IN THE XIENCE V INSTRUCTIONS FOR USE (IFU). ALTHOUGH A CONCLUSIVE CAUSE FOR THE REPORTED PATIENT EFFECTS AND THE RELATIONSHIP TO THE DEVICE, IF ANY, CANNOT BE DETERMINED, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY WITH RESPECT TO MANUFACTURE OR DESIGN. SINCE IT WAS REPORTED THAT THE XIENCE V STENT WAS IMPLANTED IN A 30 MM LONG LESION, IT SHOULD BE NOTED THAT THE XIENCE V IFU STATES: THE XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM (XIENCE V STENT) IS INDICATED FOR IMPROVING CORONARY LUMINAL DIAMETER IN PATIENTS WITH SYMPTOMATIC HEART DISEASE DUE TO DE NOVO NATIVE CORONARY ARTERY LESIONS (LENGTH LESS THAN OR EQUAL TO 28 MM) WITH REFERENCE VESSEL DIAMETERS OF 2.5 MM TO 4.25 MM. THE IFU CAUTIONS THAT: THE SAFETY AND EFFECTIVENESS OF THE XIENCE V STENT HAVE NOT BEEN ESTABLISHED FOR SUBJECT POPULATIONS WITH LESION LENGTHS GREATER THAN 28MM. IN THIS CASE, IT CANNOT BE DETERMINED WHETHER THE IFU DEVIATION CONTRIBUTED TO THE REPORTED PATIENT EFFECTS.

Description of Event or Problem · 1

IT WAS REPORTED VIA A TRIAL THAT APPROXIMATELY TWENTY MONTHS POST XIENCE V STENT IMPLANTATION IN THE MID RIGHT CORONARY ARTERY (MRCA) WITH TWO XIENCE V STENTS, THE PATIENT WAS HOSPITALIZED WITH CHEST PAIN. ECG RESULTS SHOWED NON-ST ELEVATION MYOCARDIAL INFARCTION. ON (B)(6) 2010, THE PATIENT UNDERWENT PERCUTANEOUS CORONARY INTERVENTION FOR MILD IN-STENT RESTENOSIS IN THE MID RIGHT CORONARY ARTERY. THE PATIENT'S CONDITION RESOLVED ON (B)(6) 2010. THERE WAS NO ADVERSE PATIENT SEQUELA. ALTHOUGH REQUESTED, THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 8051761

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| R| S STENT: 3.5 X 15 MM XIENCE V ((B)(4))