FDA Adverse Event Malfunction Summary report: N

REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE

MDR report key: 1964280 · Received January 19, 2011

Report

Report Number
6000001-2011-00341
Event Type
Malfunction
Date Received
January 19, 2011
Date of Event
November 16, 2010
Report Date
December 6, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED. THE REPORTED CONDITION OF BROKEN WAS CONFIRMED AS FAILURE CODE 810:11. THE ROOT CAUSE COULD NOT BE DETERMINED. NO FURTHER TESTING HAS BEEN COMPLETED AT THIS TIME AND NO REPAIRS HAVE BEEN PERFORMED AS THE REFERENCED DEVICE HAS BEEN REMOVED FROM SERVICE. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A CONDITION OF BROKEN. IT IS UNKNOWN WHEN OR AT WHAT POINT IN THE PROCESS THE REPORTED CONDITION OCCURRED. REVIEW OF THE DEVICE EVENT HISTORY CONFIRMED THIS CONDITION AS FAILURE CODE 810:11, WHICH INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1