REFURBED COLLEAGUE CX TRIPLE CCHANNEL VOLUMETRIC INFUSION PU
Report
- Report Number
- 6000001-2011-00337
- Event Type
- Malfunction
- Date Received
- January 19, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 11, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS INFORMATION WAS INCORRECT IN THE INITIAL MEDWATCH.
(B)(4). ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.
(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF FAILURE CODE 550:320:844:0000. THE ROOT CAUSE WAS DETERMINED TO BE A DISCONNECTED SPEAKER HARNESS. THE SPEAKER HARNESS WAS RECONNECTED TO REPAIR THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE BAXTER SERVICE TECHNICIAN REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED FAILURE CODE 550:320:844:0000 DURING DEVICE EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS DETERMINED TO BE A DISCONNECTED SPEAKER HARNESS, INDICATING THAT THE DEVICE FAILED TO AUDIBLY ALARM FOR THIS CONDITION. THERE WAS NO PATIENT INVOLVEMENT. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00. NO ADDITIONAL INFORMATION IS AVAILABLE.