FDA Adverse Event Death Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1964240 · Received January 19, 2011

Report

Report Number
2649622-2011-00069
Event Type
Death
Date Received
January 19, 2011
Date of Event
April 1, 2009
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY (B)(4) NO ANOMALIES FOUND, OUTER INSULATION BREACHED CUT AND COSMETIC DEPRESSION, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. LEAD MODEL 4076 - NO ANOMALIES FOUND. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY (B)(4) NO ANOMALIES FOUND, OUTER INSULATION BREACHED CUT AND COSMETIC DEPRESSION, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) - NO ANOMALIES FOUND. PROXIMAL SEGMENT RETURNED AND ANALYZED. CORRECTED SERIAL NUMBER OF (B)(4) LEAD FROM NONE TO (B)(4).

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4) NO ANOMALIES FOUND. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION BREACHED CUT AND COSMETIC DEPRESSION, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) - NO ANOMALIES FOUND. PROXIMAL SEGMENT RETURNED AND ANALYZED. CORRECTED SERIAL NUMBER OF (B)(4) LEAD FROM NONE TO (B)(4).

Description of Event or Problem · 1

THE DEVICE AND LEADS WERE RETURNED TO THE MANUFACTURER AFTER THE PATIENT'S DEATH. THE CAUSE OF DEATH AND DATE OF DEATH HAVE BEEN REQUESTED AND NOT RECEIVED. FOLLOW UP REVEALED PATIENT HAD BEEN SEEN IN CLINIC ONLY ONCE AND THAT WAS 2 MONTHS BEFORE DEATH. THE NURSE REPORTED NO DEVICE ISSUES WERE NOTED.

Description of Event or Problem · 1

THE DEVICE AND LEADS WERE RETURNED TO THE MANUFACTURER AFTER THE PATIENT'S DEATH. THE CAUSE OF DEATH AND DATE OF DEATH HAVE BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 61 YR Death