FDA Adverse Event Malfunction Summary report: N

ADVANTA

MDR report key: 1964177 · Received December 21, 2010

Report

Report Number
1824206-2010-11870
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
December 3, 2010
Report Date
December 3, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ACCOUNT STATED THEY FOUND THIS ISSUE OCCURRED WITH EIGHTEEN DIFFERENT ADVANTA BEDS THAT WERE ALL ON THE SAME FLOOR OF THEIR FACILITY. THE ACCOUNT IS NOT SURE IF THERE WAS SOME TYPE OF SURGE THAT TOOK THE UNIVERSAL TELEVISION BOARDS OUT. THE UNIVERSAL TELEVISION BOARD FAILED. THE ACCOUNT REPLACED THE UNIVERSAL TELEVISION BOARD IN ALL EIGHTEEN BEDS TO RESOLVE THE ISSUE. THE ACCOUNT DID NOT HAVE THE SERIAL NUMBERS FOR THE BEDS.

Description of Event or Problem · 1

ACCOUNT ALLEGED THAT THEY ARE JUST NOW HOOKING BEDS UP TO THE NURSE CALL IN ORDER TO SEND PRIORITY CALL WHEN PPM IS ALARMING. ACCOUNT STATED THEY HAVE FOUND MULTIPLE BEDS NOT SENDING CALLS WHEN ALARMING BUT DO ALARM LOCALLY AT THE BED. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1