FDA Adverse Event
Malfunction
Summary report: N
4 MOTOR ADVANCE BED
MDR report key: 1964142
·
Received December 21, 2010
Report
- Report Number
- 1824206-2010-11837
- Event Type
- Malfunction
- Date Received
- December 21, 2010
- Date of Event
- November 23, 2010
- Report Date
- November 23, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN INSPECTED THE AC CORD FOR DAMAGE AND FOUND NONE. HE LOWERED THE ELECTRICAL PAN AND RE-SEATED BOTH OF THE CONNECTORS AT THE POWER PCB AND THE SCREW FOR THE GROUND WIRE AT THE PAN AND HE WAS ABLE TO GET AN ACCEPTABLE READING OF 0.07 OHMS.
Description of Event or Problem · 1
DURING A PREVENTIVE MAINTENANCE CHECK THE TECHNICIAN FOUND THE GROUND RESISTANCE READING WAS OUT OF SPECIFICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 4 MOTOR ADVANCE BED | AC-POWERED ADJUSTABE HOSPITAL BED | FNL | HILL-ROM, INC. | 1105 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |