FDA Adverse Event Malfunction Summary report: N

4 MOTOR ADVANCE BED

MDR report key: 1964142 · Received December 21, 2010

Report

Report Number
1824206-2010-11837
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
November 23, 2010
Report Date
November 23, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN INSPECTED THE AC CORD FOR DAMAGE AND FOUND NONE. HE LOWERED THE ELECTRICAL PAN AND RE-SEATED BOTH OF THE CONNECTORS AT THE POWER PCB AND THE SCREW FOR THE GROUND WIRE AT THE PAN AND HE WAS ABLE TO GET AN ACCEPTABLE READING OF 0.07 OHMS.

Description of Event or Problem · 1

DURING A PREVENTIVE MAINTENANCE CHECK THE TECHNICIAN FOUND THE GROUND RESISTANCE READING WAS OUT OF SPECIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4 MOTOR ADVANCE BED AC-POWERED ADJUSTABE HOSPITAL BED FNL HILL-ROM, INC. 1105

Patients

Seq Age Sex Outcome Treatment
1