FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1964139 · Received December 21, 2010

Report

Report Number
1824206-2010-11835
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
December 10, 2010
Report Date
December 10, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THAT BRAKES WOULD NOT HOLD AND SWIVEL. THE CASTERS WERE WORN OUT. REPLACED THE BRAKE AND BRAKE/STEER CASTERS TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

INFO RECEIVED INDICATED THAT THE BRAKES WOULD NOT HOLD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900C001742-1

Patients

Seq Age Sex Outcome Treatment
1