FDA Adverse Event Malfunction Summary report: N

CCU BED

MDR report key: 1964137 · Received December 21, 2010

Report

Report Number
1824206-2010-11839
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
November 29, 2010
Report Date
November 29, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE WIRING HARNESS TO REPAIR THE BED. THIS WIRING HARNESS IS UNDER A COVER AND IS NOT EXPOSED TO THE PT OR CAREGIVER.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE SIDERAIL CONTROLS ARE NOT WORKING. THE TECHNICIAN FOUND THE WIRING HARNESS FROM THE MOTOR TO THE CIRCUIT BOARD LOOKED BURNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CCU BED AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 894

Patients

Seq Age Sex Outcome Treatment
1