FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1964113 · Received December 21, 2010

Report

Report Number
3015876-2010-01395
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
November 23, 2010
Report Date
November 23, 2010
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT WOULD NOT OPERATE ON AC OR DC POWER. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED POWER SUPPLY ASSEMBLY AT THE FAILURE ANALYSIS CENTER AND OBSERVED NO AC OPERATION. THE CAUSE OF THE MALFUNCTION WAS DETERMINED TO BE A FAILURE OF THE AC POWER SUPPLY, TRANSISTORS Q1 AND Q2 WERE SHORTED AND FUSE F1 WAS OPEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON AC OR DC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA