FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 19639547 · Received June 28, 2024

Report

Report Number
2955842-2024-16454
Event Type
Malfunction
Date Received
June 28, 2024
Date of Event
June 6, 2024
Report Date
June 6, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110744
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED PERSONALITY MODULE SURGEON CONSOLE (PMSC) BOARD TO RESOLVE THE ISSUE. THE SYSTEM WAS VERIFIED AND READY TO USE. THE UNIT WAS RETURNED FOR FAILURE ANALYSIS AND THE REPORTED FAILURE WAS REPLICATED. THIS UNIT WAS INSTALLED INTO THE TEST SYSTEM AND RUNNING VIDEO TEST, 1 MINUTE SINE CYCLE, 10 POWER CYCLES & SITTING IDLE FOR 1 HOUR. DURING THE POWER CYCLES THE TECHNICIAN REPLICATED THE REPORTED FAILURE. THE TROUBLESHOOTING FOUND THE SYSTEM CANNOT DETECTED THE LEAF WHICH LOCATED ON THE PMSC MODULE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PULMONARY LOBECTOMY SURGICAL PROCEDURE, THERE WAS NO IMAGE IN THE LEFT EYE OF THE CONSOLE. THE VISION SIDE CART (VSC) IMAGES WERE NORMAL FOR BOTH LEFT AND RIGHT SIDES. THE CUSTOMER DISCONNECTED THE DVI CABLES TO THE SURGEON SIDE CONSOLE (SSC) AND REBOOTED THE SYSTEM, BUT THE ISSUE PERSISTED. THE TECHNICAL SERVICE ENGINEER (TSE) FOUND AN ERROR 48200 IN THE LOGS POINTING TO THE PERSONALITY MODULE SURGEON CONSOLE (PMSC). THE TSE RECOMMENDED THE CUSTOMER TO CHANGE TO 2D VIEWER MODE TO CONTINUE THE PROCEDURE. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE CUSTOMER PAUSED THE PROCEDURE AND WAITED FOR THE FIELD SERVICE ENGINEER (FSE) TO FIX THE ISSUE. THERE WAS NO INJURY TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
851548 DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS NAY INTUITIVE SURGICAL, INC 380677-16 N/A 00886874110744

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose DA VINCI INSTRUMENTS AND ACCESSORIES