FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1963913
·
Received January 11, 2011
Report
- Report Number
- 2531779-2011-00091
- Event Type
- Malfunction
- Date Received
- January 11, 2011
- Date of Event
- December 8, 2010
- Report Date
- December 8, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Removal / Correction Number
- 2531779-03/24/2010/003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO REPORTED PT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED LOSS OF PRIME WARNINGS ASSOCIATED WITH ZERO FORCE AND TWO "NO CARTRIDGE DETECTED" WARNINGS. AN "EZPRIME" OPERATION WAS PERFORMED WITH NO LOSS OF PRIME WARNINGS DUPLICATED. THE PUMP DID NOT DISPENSE FLUID FROM THE CARTRIDGE DURING THE LOAD STEP.
Description of Event or Problem · 1
THE PT'S MOTHER STATED THAT THE PUMP EMITTED MULTIPLE "NO PRIME" AND "NO CARTRIDGE DETECTED" WARNINGS DURING THE LOAD STEP. THE PUMP REPORTEDLY DISPENSED ALL THE INSULIN FROM THE CARTRIDGE DURING THE LOAD STEP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 YR |