FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1963894 · Received January 11, 2011

Report

Report Number
2531779-2011-00070
Event Type
Malfunction
Date Received
January 11, 2011
Report Date
December 6, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE OK BUTTON WAS INTERMITTENTLY RESPONDING TO USER INPUTS, THE UP/DOWN/CONTRAST BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, THE CONTRAST KEY WAS INVERTED, THE UP/DOWN/OK KEYS BECAME INVERTED DURING TESTING AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UP/DOWN ARROW KEYPAD BUTTONS HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1