RENAL - DISPOSABLE
Report
- Report Number
- 1423500-2011-00765
- Event Type
- Malfunction
- Date Received
- January 19, 2011
- Date of Event
- December 25, 2010
- Report Date
- December 25, 2010
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 WAS NOT CONFIRMED DUE TO THE LACK OF A SAMPLE; HOWEVER, BASED ON THE INFORMATION OBTAINED DURING BAXTER'S INVESTIGATION, THIS INCIDENT WAS DETERMINED TO BE CAUSED BY AIR BEING SUCKED INTO THE DISPOSABLE SET AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. A BATCH REVIEW WAS NOT PERFORMED AS THE LOT INFORMATION WAS UNKNOWN. THE HOMECHOICE APD SYSTEMS PATIENT AT HOME GUIDE INSTRUCTS PATIENTS ON HOW TO DISCONNECT FOR A SHORT TIME AND RECONNECT FOR EMERGENCIES. THIS REVIEW FINDS THE LABELING FOR THE POTENTIAL USE ERROR(S) IN THIS COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4). SINCE THE PRODUCT CODE IS UNKNOWN AT THIS TIME, THERE WILL NOT BE A 510K NUMBER SUBMITTED WITH THIS REPORT.
A PATIENT CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 WHILE USING THE HOMECHOICE (HC) DURING DWELL 5 OF 6. GTS EXPLAINED THE SYSTEM ERROR INDICATED A LARGE AMOUNT OF AIR HAD ENTERED INTO THE DISPOSABLE SET, AND THEY WOULD NEED TO START OVER WITH NEW SUPPLIES. THE PATIENT ELECTED TO JUST END THERAPY. GTS THEN ADVISED THE PATIENT TO NOTIFY THEIR NURSE WITHIN 24 HOURS TO LET HER KNOW WHAT HAPPENED AND INFORM HER OF ANY MISSED THERAPY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RENAL - DISPOSABLE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE APD CYCLER |