FDA Adverse Event Malfunction Summary report: N

TEC 6 PLUS VAPORIZER

MDR report key: 1963753 · Received January 11, 2011

Report

Report Number
MW5019008
Event Type
Malfunction
Date Received
January 11, 2011
Report Date
January 11, 2011
Manufacturer
GE DATEX OHMEDA
Product Code
CAD
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
WI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

UNACCEPTABLE RATE OF FAILURE OF LEAK TEST OF ANESTHESIA EQUIPMENT. GE TEC 6 PLUS DESFLURANE VAPORIZER USED WITH A GE AVANCE ANESTHESIA MACHINE. THE AVANCE HAS A BUILT-IN SEMI-AUTOMATED LEAK TEST, CALLED THE "LOW P LEAK TEST". WE HAVE HAD FREQUENT REPEATABLE FAILURES WHEN THE DESFLURANE VAPORIZER IS INCLUDED AS PRESCRIBED IN THIS TEST, SET TO 1%. THESE FAILURES CAN CAUSE DELAYS IN SURGICAL CASES, OR NOT HAVING NEEDED EQUIPMENT AVAILABLE. WE HAVE HAD 14 VAPORIZERS FAIL AND BE REPLACED IN THE LAST 9 MONTHS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TEC 6 PLUS VAPORIZER TEC 6 PLUS VAPORIZER FOR DESFLURANE CAD GE DATEX OHMEDA

Patients

Seq Age Sex Outcome Treatment
1