FDA Adverse Event
Malfunction
Summary report: N
TEC 6 PLUS VAPORIZER
MDR report key: 1963753
·
Received January 11, 2011
Report
- Report Number
- MW5019008
- Event Type
- Malfunction
- Date Received
- January 11, 2011
- Report Date
- January 11, 2011
- Manufacturer
- GE DATEX OHMEDA
- Product Code
- CAD
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- WI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
UNACCEPTABLE RATE OF FAILURE OF LEAK TEST OF ANESTHESIA EQUIPMENT. GE TEC 6 PLUS DESFLURANE VAPORIZER USED WITH A GE AVANCE ANESTHESIA MACHINE. THE AVANCE HAS A BUILT-IN SEMI-AUTOMATED LEAK TEST, CALLED THE "LOW P LEAK TEST". WE HAVE HAD FREQUENT REPEATABLE FAILURES WHEN THE DESFLURANE VAPORIZER IS INCLUDED AS PRESCRIBED IN THIS TEST, SET TO 1%. THESE FAILURES CAN CAUSE DELAYS IN SURGICAL CASES, OR NOT HAVING NEEDED EQUIPMENT AVAILABLE. WE HAVE HAD 14 VAPORIZERS FAIL AND BE REPLACED IN THE LAST 9 MONTHS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TEC 6 PLUS VAPORIZER | TEC 6 PLUS VAPORIZER FOR DESFLURANE | CAD | GE DATEX OHMEDA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |