FDA Adverse Event Malfunction Summary report: N

MERIT CUSTOM KIT

MDR report key: 1963727 · Received December 22, 2010

Report

Report Number
1721504-2010-00459
Event Type
Malfunction
Date Received
December 22, 2010
Date of Event
October 27, 2010
Report Date
December 4, 2010
Manufacturer
MERIT MEDICAL SYSTEMS, INC.
Product Code
DQO
PMA / PMN Number
K913682
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THREE USED SUSPECT DEVICES WERE RETURNED FOR EVAL. THE COMPLAINT WAS CONFIRMED. THE ROTATOR SEPARATED AT THE BOND BETWEEN THE COLLAR AND THE HOUSING. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT REVEAL ANY EXCEPTION DOCUMENTS. COMPLAINT DATA BASE WAS REVIEWED AND FOUND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER. THE ROOT CAUSES OF THE FAILURE ARE ATTRIBUTED TO THE MFG BONDING PROCESS. CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED TO ADDRESS THIS TYPE OF FAILURE. THIS LOT WAS BUILT PRIOR TO IMPLEMENTATION OF THE NOTED CORRECTIVE ACTIONS. COMPLAINT DATA WILL BE MONITORED TO VERIFY EFFECTIVENESS OF CORRECTIVE ACTIONS TAKEN. EVAL METHOD: DEVICE HISTORY RECORD WAS REVIEWED, COMPLAINT DATA BASE WAS REVIEWED.

Description of Event or Problem · 1

THE ROTATOR BROKE DURING A LEFT VENTRICULOGRAM PROCEDURE THREE TIMES IN A ROW. PRESSURE LIMIT 900 PSI. NO HARM OR INJURY WAS REPORTED. THIS IS ONE OF THREE REPORTS FOR THIS COMPLAINT: 1721504-2010-00457, 1721504-2010-00458.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MERIT CUSTOM KIT CATHETER, INTRAVASCULAR, DIAGNOSITC DQO MERIT MEDICAL SYSTEMS, INC. F627770

Patients

Seq Age Sex Outcome Treatment
1 CONTRAST MEDIA