ID NOW COVID-19 2.0 TEST KIT 24T (EUA)
Report
- Report Number
- 1221359-2024-00433
- Event Type
- Malfunction
- Date Received
- June 28, 2024
- Date of Event
- June 10, 2024
- Report Date
- July 8, 2024
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QWR
- UDI-DI
- 00811877011354
- PMA / PMN Number
- K221925
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D2A - COMMON DEVICE NAME: SIMPLE POINT-OF-CARE DEVICE TO DETECT SAR-COV-2 NUCLEIC ACID TARGETS FROM CLINICAL SPECIMENS IN NEAR-PATIENT SETTINGS TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M821209 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 192-000/LOT M821209 AND TEST BASE PART NUMBER 192-430/LOT M821209. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M821209 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE, HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE COULD BE PATIENT SAMPLE INTERFERENCE.
D2A - COMMON DEVICE NAME: SIMPLE POINT-OF-CARE DEVICE TO DETECT SAR-COV-2 NUCLEIC ACID TARGETS FROM CLINICAL SPECIMENS IN NEAR-PATIENT SETTINGS THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION OR UPON RECEIPT OF NEW INFORMATION.
THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2024 ON A NASAL SAMPLE USING THE KIT SWAB. THIS MANUFACTURER¿S REPORT ADDRESSES TEST TWO (2) OF TWO (2) FOR PATIENT THREE (3) OF THREE (3). THE CUSTOMER PERFORMED AN ADDITIONAL TEST ON THE SAME DAY ON A DIFFERENT ID NOW INSTRUMENT AND GENERATED A NEGATIVE RESULT. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT¿S TREATMENT.
THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2024 ON A NASAL SAMPLE USING THE KIT SWAB. THIS MANUFACTURER¿S REPORT ADDRESSES TEST TWO (2) OF TWO (2) FOR PATIENT THREE (3) OF THREE (3). THE CUSTOMER PERFORMED AN ADDITIONAL TEST ON THE SAME DAY ON A DIFFERENT ID NOW INSTRUMENT AND GENERATED A NEGATIVE RESULT. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT¿S TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2192504 | ID NOW COVID-19 2.0 TEST KIT 24T (EUA) | SIMPLE POINT-OF-CARE DEVICE TO DETECT SAR-COV-2 NUCLEIC ACID TARGETS | QWR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | M821209 | 00811877011354 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |