FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 1963608
·
Received December 20, 2010
Report
- Report Number
- 8020893-2010-00586
- Event Type
- Malfunction
- Date Received
- December 20, 2010
- Date of Event
- December 5, 2010
- Report Date
- December 7, 2010
- Manufacturer
- PURITAN BENNETT CORP.
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE SERVICE REPORT SHOWS, THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE PURITAN BENNETT CUSTOMER SERVICE ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED BDU CPU, GUI CPU, AND BACKLIGHT PCB. THE UNIT PASSED EXTENDED SELF TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 840 VENTILATOR | VENTILATOR | CBK | PURITAN BENNETT CORP. | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |