FDA Adverse Event Malfunction Summary report: N

STENOSCOPE

MDR report key: 1963558 · Received December 20, 2010

Report

Report Number
9617766-2010-00864
Event Type
Malfunction
Date Received
December 20, 2010
Date of Event
December 8, 2010
Report Date
December 20, 2010
Manufacturer
GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. A CABLE WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM BECAME UNABLE TO PRODUCE FLUOROSCOPY XRAY. NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STENOSCOPE FLUOROSCOPE XRAY JAA GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE) STENOSCOPE

Patients

Seq Age Sex Outcome Treatment
1