SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA
Report
- Report Number
- 2023826-2011-00037
- Event Type
- Malfunction
- Date Received
- January 12, 2011
- Date of Event
- December 21, 2010
- Report Date
- December 23, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P900048
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). EVAL: RESULTS (OTHER): VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED A PIECE OF THE OPTIC AND A HAPTIC WERE TORN OFF AND MISSING AND THERE WAS A DARK SURGICAL RESIDUE ON THE SURFACE OF THE LENS. CONCLUSION: (OTHER): AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN (B)(4) 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MFG OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVAL OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).
IT WAS REPORTED THAT THE SURGEON ATTEMPTED TO INSERT AN AQ2015A THREE PIECE SILICONE LENS AND THE LENS TORE WHILE ADVANCING TOWARDS THE EYE. NO PT CONTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | AQ2015A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE: MODEL AQ CARTRIDGE-FP, LOT # UNK| INJECTOR: MODEL MSI-TM, LOT # UNK |