FDA Adverse Event Malfunction Summary report: N

ACTIVA RC

MDR report key: 1963492 · Received December 20, 2010

Report

Report Number
3007566237-2010-10604
Event Type
Malfunction
Date Received
December 20, 2010
Date of Event
December 7, 2010
Report Date
December 7, 2010
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
MHY
PMA / PMN Number
P960009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

RECEIVED INFO FROM HCP THAT POST-OPERATIVELY WHEN DEVICE WAS INTERROGATED, BOTH TIMES WERE HIGH IMPEDANCES >4000. THIS WAS NOTED ON ALL COMBOS WITH #8, C-8=4258, 8-9=3483, 8-10=4677 AND 8-10=4677, ALL OTHER COMBOS WERE NORMAL. IMPEDANCES WERE NOT CHECKED INTRA-OPERATIVELY BECAUSE PT HAD NORMAL IMPEDANCES AND GOOD THERAPY GOING INTO THE PROCEDURE AND IT WAS JUST AN INS CHANGE OUT. PATIENT'S PREVIOUS THERAPY WAS PROGRAMMED USING THE #8 CONTACT, C+8-, 4,9V, 180US 60HZ (DYSTONIA). HCP IS UNSURE IF OTHER PROGRAMMING WOULD BE EFFECTIVE FOR THE PT IF #8 IS NOT AVAILABLE. ADDITIONAL INFO REPORTS PT AND HIS WIFE HAD SOME RECHARGING ISSUES, BUT THIS WAS FELT TO BE CAUSED BY INCISIONAL SWELLING AND THE BANDAGES PRESENT. ADDITIONAL INFO HAS BEEN REQUESTED AND IF RECEIVED, A FOLLOW UP REPORT WILL BE SENT. REFERENCE MFR REPORT # 3004209178201010605.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACTIVA RC MHY MEDTRONIC NEUROMODULATION 37612 NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR IMPLANTABLE NEURO STIMULATOR: MODEL 37612| UNK CONVERSION TYPE: MODEL 3550-09, LOT# N170731| IMPLANTABLE NEURO STIMULATOR: MODEL 7428| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| LOT# NFD111031H| EXPLANTED:| STIM ACCESSORY: MODEL 37092, LOT# 240620002| UNK CONVERSION TYPE: MODEL 3550-09, LOT# N166552| IMPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| IMPLANTED:| ADAPTER: MODEL 64001, LOT# N259554| EXPLANTED:| LOT# NKG701456H| IMPLANTED:| EXPLANTED:| EXTENSION: MODEL 7482, LOT# NHU060805V| EXPLANTED:| EXTENSION: MODEL 7482, LOT# NHU104954V| EXPLANTED:| EXPLANTED:| EXPLANTED:| LEAD: MODEL 3387, LOT# J0428156V| EXPLANTED:| IMPLANTED:| LOT# NFD111032H| IMPLANTABLE NEURO STIMULATOR: MODEL 7428| ACCESSORY: MODEL 37651, LOT# NKA146010N| LEAD: MODEL 3387, LOT# J0428156V| IMPLANTED:| LEAD: MODEL 3387, LOT# J0437506V| IMPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 37642, LOT# NJZ108241N| IMPLANTED:| EXPLANTED: