COOK MEDICAL PNEUMOTHORAX SET
MDR report key: 1963476
·
Received January 14, 2011
Report
- Report Number
- 1963476
- Event Type
- Injury
- Date Received
- January 14, 2011
- Report Date
- January 11, 2011
- Manufacturer
- COOK INCORPORATED
- Product Code
- BYY
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
CHEST TUBE PLACED ON (B)(6) 2011 FOR PNEUMO THORAX (AFTER LUNG BIOPSY ON (B)(6)-2011) - PROCEDURE PERFORMED BY IR PHYSICIAN; THE HEIMLICH VALVE WAS PLACED ON THE CHEST TUBE; POST-PROCEDURE, SCAN DEMONSTRATES IMPROVEMENT IN PNEUMOTHORAX. THE PT DEVELOPED SUBCUTANEOUS AIR ON RIGHT CHEST TO RADIOLOGY - CT PERFORMED & NOTED LARGE PNEUMOTHORAX - AT TIME PT SEEN IN RADIOLOGY, THE VALVE ON THE TUBING WAS CONNECTED BACKWARDS - UNCLEAR HOW THIS OCCURRED - UPON CLOSE INSPECTION, THERE IS A CLEAR ARROW DENOTING, DIRECTION OF FLOW BUT THIS INFO IS DIFFICULT TO SEE & READ; IF THE TUBING & VALVE CAME APART. IT WOULD BE DIFFICULT TO KNOW THAT THE VALVE WAS TO BE PLACED IN A SPECIFIC DIRECTION; ASSURES THAT THIS IS WHAT HAPPENED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COOK MEDICAL PNEUMOTHORAX SET | VALVE FOR CHEST TUBE SET | BYY | COOK INCORPORATED | C-TPT-100 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Hospitalization| R |