FDA Adverse Event Malfunction Summary report: N

HARMONY IQ

MDR report key: 19634127 · Received June 28, 2024

Report

Report Number
3008776287-2024-00011
Event Type
Malfunction
Date Received
June 28, 2024
Date of Event
June 11, 2024
Report Date
June 28, 2024
Manufacturer
BLACK DIAMOND VIDEO, INC.
Product Code
FTA
UDI-DI
00724995199142
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

STERIS IS ACTIVELY WORKING WITH THE CUSTOMER AND EVALUATING THE REPORTED ISSUE ONSITE AND REMOTELY. STERIS HAS PROPOSED A PLAN WITH THE CUSTOMER AND ORDERED PARTS TO INSTALL TO MONITOR AND ADDRESS THIS ISSUE. OUR INVESTIGATION INTO THE ROOT CAUSE OF REPORTED EVENT IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

IN JULY 2024 AS ANOTHER SOLUTION TO MITIGATE THE ISSUE, AN EDID MINDER WAS INSTALLED IN OPERATING ROOM 8. THE EDID MINDER MONITORS THE DISPLAY AND COMBINES THE DISPLAY DATA WITH ITS INTERNAL EDID TABLE TO SEND BACK TO THE HIQ SYSTEM ALLOWING FOR MOST STABLE COMMUNICATION BETWEEN THE RECEIVE MODEM AND THE DISPLAY. DURING THIS VISIT, STERIS WITH THE SUPPORT OF THE SOFTWARE DEVELOPER ATTEMPTED TO RECREATE THE PROBLEM. THE OPERATING ROOM WAS AVAILABLE FOR ONE HOUR AND WE WERE UNABLE TO RECREATE THE ISSUE. SINCE INSTALLATION OF THE EDID MINDER, THERE HAVE BEEN NO ISSUES REPORTED TO STERIS BY THE CUSTOMER IN THIS OPERATING ROOM. STERIS WILL INSTALL AND UTILIZE MONITORING EQUIPMENT IN OPERATING ROOM6 AND OPERATING ROOM 8 TO SEE IF THE ISSUE RECURS AND DETERMINE A ROOT CAUSE. IF THE ISSUE DOES NOT RECUR IN OPERATING ROOM 6 AND OPERATING ROOM 8 AND FURTHER TESTING IS INCONCLUSIVE, THEN SIMILAR HARDWARE TO THE EDID MINDER MAY BE INSTALLED IN THE REMAINING OPERATING ROOMS. STERIS WILL COMPLETE ANOTHER ONSITE EXTENSIVE TROUBLESHOOTING SESSION TO ATTEMPT TO RECREATE THE ISSUE IN ONE OF THE REMAINING OPERATING ROOMS THE WEEK OF SEPTEMBER 16 AND COLLECT DATA TO DETERMINE ROOT CAUSE.

Additional Manufacturer Narrative · 0

THE LIMA SCALERS HAVE BEEN INSTALLED AT THE FACILITY WHERE APPROPRIATE. STERIS HAS CONTINUED TO MONITOR THE PERFORMANCE OF THE DEVICES WITH THE CUSTOMER SINCE THE INSTALLATION OF THE LIMA SCALERS. TO DATE, THERE HAVE BEEN NO RECURRENCES OF THE REPORTED ISSUE WITH THE LIMA SCALERS INSTALLED. AT THIS TIME, NO FURTHER ACTION WILL BE TAKEN AS NO ADDITIONAL ISSUES HAVE BEEN REPORTED. STERIS WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS TO ENSURE THE PRODUCT CONTINUES TO PERFORM AS EXPECTED.

Additional Manufacturer Narrative · 0

ON (B)(6) 2024, STERIS SERVICE AND ENGINEERING ARRIVED ONSITE TO PERFORM TESTING THROUGHOUT THE WEEK IN THE EVENING WHILE THE ORS WERE AVAILABLE. THE BLINKING ISSUE WAS DUPLICATED ONLY WHILE ROUTING A STORZ CAMERA (50 HZ VIDEO) THROUGH THE CAPTURE CARD. ADDITIONAL TESTING WAS PERFORMED TO VERIFY THE ISSUE WAS ONLY PRESENT WITH 50HZ VIDEO THROUGH THE CAPTURE CARD. TO RESOLVE THE ISSUE, A LIMA SCALER WAS INSTALLED. THE LIMA SCALER CONVERTS THE 50HZ TIMING FROM THE STORZ CAMERA TO 60HZ TO BE COMPATIBLE WITH THE CAPTURE CARD. ADDITIONAL TESTING AND MONITORING WERE PERFORMED FOLLOWING INSTALLATION OF THE LIMA SCALER TO ENSURE THAT THE INTERMITTENT ISSUE DID NOT RECUR. BASED ON THE RESULTS OF OUR IN-DEPTH ONSITE INSPECTION, THE ISSUE IS RELATED TO HOW THE CAPTURE CARD HANDLED THE 50HZ VIDEO. ADDITIONAL EQUIPMENT (LIMA SCALERS) WILL BE INSTALLED ON ANY 50HZ VIDEO SOURCES WHICH CANNOT BE CHANGED TO 60HZ. STERIS WILL CONTINUE TO MONITOR TO ENSURE THAT THE ISSUE DOES NOT RECUR FOLLOWING INSTALLATION OF THE LIMA SCALERS.

Description of Event or Problem · 0

THE USER FACILITY REPORTED THAT THEIR HARMONY IQ INTEGRATION SYSTEM WAS FROZEN AND HAD TO BE RESTARTED. NO REPORT OF INJURY OR PROCEDURE DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
828948 HARMONY IQ INTEGRATION SYSTEM FTA BLACK DIAMOND VIDEO, INC. HARMONY IQ 3600 00724995199142

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown