FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT XCELERANT-HYDRO

MDR report key: 1963318 · Received January 14, 2011

Report

Report Number
2953200-2011-00119
Event Type
Injury
Date Received
January 14, 2011
Date of Event
January 1, 2011
Report Date
November 9, 2012
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION - RESULTS: (ENDOLEAK), OTHER (SOURCE OF ENDOLEAK IS UNK). CONCLUSION: OTHER (SOURCE OF ENDOLEAK IS UNK).

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT HAD ANEURYSMAL GROWTH OF 8MM AT THE 2 YEAR FOLLOW UP. EXACT EVENT DATE IS UNKNOWN.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.9 CM ABDOMINAL AORTIC ANEURYSM. AT IMPLANT THE ANEURYSM WAS FUSIFORM WITH A 12 DEGREE INFRARENAL ANGLE; 31 MM AORTIC NECK. THERE WAS A LEFT ILIAC ANEURYSM WITH A 24 MM DIAMETER X 25 MM LENGTH. THE VESSEL MORPHOLOGY IS REPORTED AS HAVING MILD BILATERAL ILIAC TORTUOSITY. A TYPE IV ENDOLEAK WAS OBSERVED AT THE END OF THE PROCEDURE. THIS WAS NOT CORRECTED. A TYPE IIA, POSTERIOR, LUMBAR ORIGIN ENDOLEAK WAS NOTED AS A NEW EVENT ON THE CT WITH CONTRAST AT THE 1 MONTH VISIT. NO TYPE IV ENDOLEAK WAS NOTED. AT THE 12 MONTH REVIEW, THE ULTRASOUND SHOWED AN ENDOLEAK, TYPE UNDETERMINED, THAT WAS NOTED AS HAVING BEEN SEEN ON THE LAST SCAN DATE. NO ADDITIONAL DETAILS HAVE BEEN OBTAINED. NO CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED: IT WAS REPORTED THAT THERE IS 1 MM OF ANEURYSM GROWTH NOTED IN THE 3 YEAR SITE IMAGE REPORT. THE ANEURYSM AT TWO YEAR FOLLOW UP WAS 6.5 CM IN DIAMETER AND CURRENTLY THE DIAMETER IS 6.6 CM IN DIAMETER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT XCELERANT-HYDRO SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00386356

Patients

Seq Age Sex Outcome Treatment
1 81 YR Required Intervention