OT PING METER
Report
- Report Number
- 2939301-2011-00578
- Event Type
- Malfunction
- Date Received
- January 18, 2011
- Report Date
- January 9, 2011
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K) # IS K080639.
THE LAY USER/PATIENT'S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE HAS PASSED TESTING WITH NO FAULTS FOUND. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.
ON (B)(6) 2011, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HER ONETOUCH PING METER READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON (B)(6) 2010 AT 5PM. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "124 MG/DL" WITH THE SUBJECT METER AND "40 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 30% AND/OR <= 30 MG/DL. THE CCA WAS ADVISED THE PATIENT MANAGES HER DIABETES WITH INSULIN (TYPE/ AMOUNT NOT SPECIFIED). IT IS NOT KNOWN WHAT ACTION THE PATIENT TOOK AT THE TIME OF THE ALLEGED ISSUE. A FEW MINUTES AFTER THE ALLEGED ISSUE BEGUN, THE PATIENT CLAIMED SHE FELT A SYMPTOM OF DIZZY. THE PATIENT TOOK GLUCOSE TABLET/ GLUCOSE GEL AS TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT AND AN APPROVED SAMPLE SITE WAS USED FOR TESTING. THE PATIENT DID NOT HAVE CONTROL SOLUTION TO TEST THE SUBJECT METER. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THERE IS NO INDICATION THAT THE REPORTED ISSUE CAUSED OR CONTRIBUTED TO A SERIOUS INJURY SINCE THE PATIENT SYMPTOMS DID NOT CORRELATE WITH LFS'S DEFINITION SUGGESTIVE OF SEVERE HYPOGLYCEMIA OR HYPERGLYCEMIA. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE SUBJECT METER DID NOT MEET LFS' ACCURACY CRITERIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 2999835 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |