FDA Adverse Event
Malfunction
Summary report: N
SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER
MDR report key: 1963111
·
Received January 18, 2011
Report
- Report Number
- 2050012-2011-00137
- Event Type
- Malfunction
- Date Received
- January 18, 2011
- Date of Event
- December 21, 2010
- Report Date
- December 21, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K881495
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND FOUND A SECOND QUAD RING ON THE CO2 REFERENCE ELECTRODE CAUSING IMPROPER SEAL. THE FSE REMOVED ALL OF THE ELECTRODES, CLEANED THE FLOW CELL, AND REINSTALLED ALL ELECTRODES PROPERLY. THE FSE ALSO PERFORMED THE NECESSARY ALIGNMENTS ON THE SAMPLE WHEEL THAT WERE MISALIGNED. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES AND ALL RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT CARBON DIOXIDE (CO2) WAS LEAKING FROM THE CO2 MEASURING ELECTRODE WHERE IT THREADS INTO THE FLOW CELL IN THE SYNCHRON CX5 DELTA CHEMISTRY ANALYZER. NO OPERATOR EXPOSURE WAS NOTED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | CX5 DELTA | N/A |