FDA Adverse Event Malfunction Summary report: N

SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER

MDR report key: 1963111 · Received January 18, 2011

Report

Report Number
2050012-2011-00137
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
December 21, 2010
Report Date
December 21, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K881495
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND FOUND A SECOND QUAD RING ON THE CO2 REFERENCE ELECTRODE CAUSING IMPROPER SEAL. THE FSE REMOVED ALL OF THE ELECTRODES, CLEANED THE FLOW CELL, AND REINSTALLED ALL ELECTRODES PROPERLY. THE FSE ALSO PERFORMED THE NECESSARY ALIGNMENTS ON THE SAMPLE WHEEL THAT WERE MISALIGNED. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES AND ALL RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT CARBON DIOXIDE (CO2) WAS LEAKING FROM THE CO2 MEASURING ELECTRODE WHERE IT THREADS INTO THE FLOW CELL IN THE SYNCHRON CX5 DELTA CHEMISTRY ANALYZER. NO OPERATOR EXPOSURE WAS NOTED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. CX5 DELTA N/A