FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1963107 · Received January 18, 2011

Report

Report Number
1423500-2011-00750
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
January 4, 2011
Report Date
January 4, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE LOT NUMBER WAS NOT PROVIDED; THEREFORE A BATCH REVIEW CANNOT BE CONDUCTED. THE SAMPLE WAS DISCARDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT FOR A LOW DRAIN VOLUME ALARM (LDVA) THAT OCCURRED DURING INITIAL DRAIN WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. NO ROOT-CAUSE HAS BEEN DETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR A LOW DRAIN VOLUME ALARM, THE CARE GIVER (CG) STATED THERE WAS AIR IN THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE CG CYCLE POWER, END THERAPY AND RECOMMEND STARTING OVER WITH NEW SUPPLIES. PRODUCT SURVEILLANCE CONTACTED THE CG REGARDING THE REPORTED PROBLEM. THE CG VERIFIED THAT THERE WERE NO LOOSE CONNECTIONS, DISCONNECTIONS OR DEFECTS ON THE SUPPLIES. PER CG, THE HOME PATIENT (HP) DID NOT RECEIVE ANY INJURY OR MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT. THE CG STATED THAT THE HP WAS DOING FINE AND CONTINUING THERAPY WITHOUT ISSUES. THE CG STATED THAT SHE HAD DISCARDED THE SUPPLIES AFTER THERAPY, AND DID NOT KNOW THE LOT NUMBERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 70 YR HOMECHOICE CYCLER