FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1962718 · Received January 11, 2011

Report

Report Number
2531779-2011-00175
Event Type
Injury
Date Received
January 11, 2011
Date of Event
December 13, 2010
Report Date
December 13, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PT REPORTED THAT HE ADMINISTERED EXCESS INSULIN BY PUSHING ON THE CARTRIDGE PLUNGER WHILE ATTACHED TO THE INFUSION SET. THE USER GUIDE CAUTIONS THE USER TO FOLLOW INSTRUCTIONS IN ORDER TO AVOID IMPROPER OR INCORRECT USE OF THE PUMP. USING THE CARTRIDGE OUTSIDE OF THE PUMP IS NOT A RECOMMENDED METHOD OF INSULIN DELIVERY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PT EXPERIENCED HYPOGLYCEMIA THAT WAS TREATED BY PARAMEDICS. THE PT SAID THAT HIS WIFE CALLED FOR PARAMEDICS BECAUSE HIS BLOOD GLUCOSE READING DISPLAYED "LOW" ON THE METER. HE SAID THAT THE PARAMEDICS REMOVED THE PUMP AND HIS BLOOD GLUCOSE RESOLVED TO 200 MG/DL AFTER TREATMENT. THE PT NOTED THE FOLLOWING SEQUENCE OF EVENTS: HE ATE DINNER AROUND 7:30 PM, THE CARTRIDGE NEEDED TO BE REPLACED, HE REMOVED THE CARTRIDGE FROM THE PUMP, AND MANUALLY PUSHED ON THE PLUNGER TO DELIVER HIS DINNER BOLUS. HE SAID HE WAS NOT AWARE OF HOW MANY UNITS WERE DELIVERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. IR 1200/1250/2020/OTP

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention