HOMECHOICE CYCLER-REFURBISHED
Report
- Report Number
- 1423500-2011-00718
- Event Type
- Malfunction
- Date Received
- January 18, 2011
- Date of Event
- December 29, 2010
- Report Date
- December 29, 2010
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED AND EVALUATED. THE DEVICE FAILED THE HOMECHOICE RITE (RETURN INSTRUMENT TEST / EVALUATION) FUNCTIONAL TEST FOR VOLUMETRIC ACCURACY BUT PASSED THE RITE ELECTRICAL TEST. THE PAL EVALUATED THE DEVICE. THE ASSIGNABLE CAUSE FOR THE RITE TEST FAILURE - VOLUMETRIC ACCURACY WAS UNDETERMINED. THE DEVICE'S SERVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE RITE FAILURE. THE DEVICE DID NOT MEET SPECIFICATIONS. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED FLUID VOLUME ACCURACY TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE CYCLER-REFURBISHED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |