FDA Adverse Event
Malfunction
Summary report: N
ESPRIT VENTILATOR
MDR report key: 1962558
·
Received January 18, 2011
Report
- Report Number
- 2031642-2011-00013
- Event Type
- Malfunction
- Date Received
- January 18, 2011
- Date of Event
- December 21, 2010
- Report Date
- December 21, 2010
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- PMA / PMN Number
- K981072
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
POWER SUPPLY.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE VENTILATOR WOULD NOT OPERATE ON AC POWER, ONLY ON BACKUP BATTERY POWER. THE MANUFACTURER'S SERVICE TECHNICIAN WAS ABLE TO DUPLICATE THE REPORTED PROBLEM. THE SERVICE TECHNICIAN REPLACED THE POWER SUPPLY TO CORRECT THE REPORTED PROBLEM. APPLICABLE FINAL TESTING WAS PERFORMED AND TESTS PASSED TO OPERATING SPECIFICATIONS. IF A FAILURE OF THE POWER SUPPLY OCCURS DURING USE AND THE VENTILATOR SHUTS DOWN DUE TO A LOSS OF AC POWER, AN AUDIBLE POWER FAIL ALARM WILL ACTIVATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ESPRIT VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V1000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |