FDA Adverse Event Injury Summary report: N

NOVASURE IMPEDANCE CONTROLLED EA SYSTEM

MDR report key: 1962519 · Received January 7, 2011

Report

Report Number
1222780-2011-00005
Event Type
Injury
Date Received
January 7, 2011
Date of Event
September 1, 2010
Report Date
December 10, 2010
Manufacturer
HOLOGIC
Product Code
MNB
PMA / PMN Number
P010013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

LOT AND SERIAL NUMBER OF THE DISPOSABLE DEVICE NOT PROVIDED BY THE COMPLAINANT, THEREFORE THE EXPIRATION DATE IS NOT KNOWN. SERIAL NUMBER OF THE RADIO FREQUENCY CONTROLLER NOT PROVIDED BY THE COMPLAINANT. THE DEVICE IS NOT BEING RETURNED THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. LOT NUMBER NOT PROVIDED BY THE COMPLAINANT, THEREFORE THE MANUFACTURE DATE IS NOT KNOWN. BASED ON THE INFO OBTAINED TO DATE, NO DIRECT CORRELATION CAN BE MADE BETWEEN THE REPORTED EVENT AND THE NOVASURE SYSTEM. DEVICE HISTORY RECORD (DHR) REVIEW COULD NOT BE CONDUCTED FOR THE DISPOSABLE DEVICE AS LOT NUMBER WAS NOT PROVIDED BY THE COMPLAINANT. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ATTEMPTED NOVASURE ENDOMETRIAL ABLATION PROCEDURE ON PT THAT WAS "ONLY 8 WEEKS POST PARTUM", THE PHYSICIAN INSERTED THE DEVICE AND "REALIZED HE PERFORATED THE UTERUS." THE PROCEDURE WAS ABANDONED AND A PERFORATION WAS CONFIRMED POST HYSTEROSCOPY. THE PT WAS DISCHARGED HOME. ON (B)(6) 2010, THE PHYSICIAN REPORTED HE HAS SEEN THE PT ON F/U AND SHE WAS "FINE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NOVASURE IMPEDANCE CONTROLLED EA SYSTEM MNB HOLOGIC NS2000 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other RADIO FREQUENCY CONTROLLER - SERIAL NUMBER UNK