FDA Adverse Event Malfunction Summary report: N

PROXIMATE** RELOADABLE VASCULAR LINEAR STAPLER - 30 MM

MDR report key: 1962472 · Received January 18, 2011

Report

Report Number
3005075853-2011-00234
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
December 16, 2010
Report Date
December 17, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION WAS REQUESTED. THE ANALYSIS RESULTS SHOWED THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED ON THE DEVICE. THE RELOAD WAS RECEIVED FULLY LOADED WITH STAPLES. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED RELOAD AND IT FIRED AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLES WERE NOTED TO HAVE A PROPER B-FORMATION AND THE STAPLE LINE WAS COMPLETE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS USED DURING AN UNKNOWN PROCEDURE. THE DEVICE COULD NOT BE FIRED AS NORMAL. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE** RELOADABLE VASCULAR LINEAR STAPLER - 30 MM STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA F4PN4N

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE