PROXIMATE** RELOADABLE VASCULAR LINEAR STAPLER - 30 MM
Report
- Report Number
- 3005075853-2011-00234
- Event Type
- Malfunction
- Date Received
- January 18, 2011
- Date of Event
- December 16, 2010
- Report Date
- December 17, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION WAS REQUESTED. THE ANALYSIS RESULTS SHOWED THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED ON THE DEVICE. THE RELOAD WAS RECEIVED FULLY LOADED WITH STAPLES. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED RELOAD AND IT FIRED AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLES WERE NOTED TO HAVE A PROPER B-FORMATION AND THE STAPLE LINE WAS COMPLETE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT THE DEVICE WAS USED DURING AN UNKNOWN PROCEDURE. THE DEVICE COULD NOT BE FIRED AS NORMAL. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE** RELOADABLE VASCULAR LINEAR STAPLER - 30 MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | F4PN4N |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE |