FDA Adverse Event Injury Summary report: N

LINER: MPACT DM DM CONVERTER TIN COATED D/DMB

MDR report key: 19623329 · Received June 27, 2024

Report

Report Number
3005180920-2024-00440
Event Type
Injury
Date Received
June 27, 2024
Date of Event
May 28, 2024
Report Date
June 27, 2024
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630971256347
PMA / PMN Number
K211891
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 03 JUNE 2024 LOT 2302176: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-JUL-2023. EXPIRATION DATE: 2028-06-24. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL COMPONENTS INVOLVED IN THE EVENT: LINER: MPACT DM 01.26.2244MHC DOUBLE MOBILITY HC LINER 22.2/DMB (K131458) LOT 2314368: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 28-SEP-2023. EXPIRATION DATE: 2028-09-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

THE PATIENT HAD A PRIMARY WITH MPACT NO HOLE CUP, MPACT DM CONVERTER AND DOUBLE MOBILITY LINER IMPLANTED ON THE (B)(6) 2024. THE CUP WAS PLACED AS USUAL, 40 INCLINATION AND ROUGHLY 20 ANTEVERSION. THE PATIENT SHOWED SIGNS OF A SPINAL DEFORMITY. THE PATIENT DISLOCATED POSTERIORLY 3 TIMES OVER IN 2 WEEKS. IT WAS FOUND THAT THE PATIENT HAD AN EXTREMELY HYPER MOBILE HIP (EXTREMELY LAX IN HER LIGAMENTS) AND HER HIP POSTERIOR CAPSULE WAS DETACHED. THE SURGEON DECIDED TO REMOVE THE MPACT NO HOLE CUP WITH DM CONVERTER LINER, THE DOUBLE MOBILITY LINER AND PUT IN A HIGHLY CONSTRAINED CUP (OF COMPETITOR). THIS WAS DONE ON THE (B)(6) 2024. THE SURGEON REPORTED THAT HE DOES NOT BELIEVE THIS WAS AN IMPLANT FAULT, THE PATIENT JUST REQUIRED A MORE CONSTRAINT HIP CONFIGURATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434907 LINER: MPACT DM DM CONVERTER TIN COATED D/DMB HIP DM CONVERTER TIN COATED LZO MEDACTA INTERNATIONAL SA 2302176 07630971256347

Patients

Seq Age Sex Outcome Treatment
1 75 YR Female Required Intervention