CE INFUSOR LV 5, 12 PACK
Report
- Report Number
- 6000001-2011-00319
- Event Type
- Malfunction
- Date Received
- January 17, 2011
- Date of Event
- December 23, 2010
- Report Date
- December 23, 2010
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K041738
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). DEVICE EVALUATION: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. A VISUAL EVALUATION WAS PERFORMED AND CONFIRMED THE REPORTED CONDITION OF A LEAK. THIS DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. THE ROOT CAUSE WAS DETERMINED TO BE SEPARATED TUBING AT THE JUNCTION OF THE LUER. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. UPON COMPLETION OF BAXTER?S INVESTIGATION A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).
THIS IS A REPORT FROM BAXTER (B)(6) OF AN LV5 INFUSOR THAT WAS LEAKING. THE INFUSOR WAS FILLED WITH FLUOROURACIL AT THE DUBLIN COMPOUNDING. NO PATIENT WAS INVOLVED. NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THIS DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CE INFUSOR LV 5, 12 PACK | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 10J060 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FLUOROURACIL |