FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL

MDR report key: 1962194 · Received January 17, 2011

Report

Report Number
6000001-2011-00318
Event Type
Malfunction
Date Received
January 17, 2011
Date of Event
October 1, 2010
Report Date
December 23, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A TREND REVIEW HAS BEEN PERFORMED AND THERE HAVE BEEN SIMILAR REPORTS MADE TO BAXTER. THE ROOT CAUSE WILL CONTINUE TO BE INVESTIGATED THROUGH SIMILAR TRENDS.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE CAUSE WAS A FAULTY PHM (PUMPHEAD MODULE). THE PHM HAS BEEN REPLACED. ADDITIONAL: A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE HAS NOT BEEN PREVIOUSLY SERVICED FOR THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4).THE DEVICE HAS BEEN RECEIVED AND IS CURRENTLY BEING EVALUATED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WITH THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

DURING BAXTER'S REVIEW OF THE EVENT HISTORY FOR ANOTHER REPORT, IT WAS DISCOVERED THAT FAILURE CODE 814:02 OCCURRED ONCE DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 6.13.90, CATEGORIZED AS REMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1