COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL
Report
- Report Number
- 6000001-2011-00317
- Event Type
- Malfunction
- Date Received
- January 17, 2011
- Date of Event
- October 8, 2010
- Report Date
- December 23, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE CAUSE WAS A FAULTY PHM (PUMPHEAD MODULE). THE PHM HAS BEEN REPLACED. ADDITIONAL: A SERVICE HISTORY REVIEW HAS BEEN PERFORMED AND THE DEVICE HAS NOT BEEN PREVIOUSLY SERVICED FOR THE REPORTED CONDITION.
(B)(4). THE DEVICE HAS BEEN RECEIVED AND IS CURRENTLY BEING EVALUATED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WITH THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAIALBLE.
(B)(4). A TREND REVIEW HAS BEEN PERFORMED AND THERE HAVE BEEN SIMILAR REPORTS MADE TO BAXTER. THE ROOT CAUSE WILL CONTINUE TO BE INVESTIGATED THROUGH SIMILAR TRENDS.
DURING BAXTER'S REVIEW OF THE EVENT HISTORY FOR ANOTHER REPORT, IT WAS DISCOVERED THAT FAILURE CODE 814:02 OCCURRED ONCE DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 6.13.90, CATEGORIZED AS REMEDIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |