FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 1962169 · Received January 17, 2011

Report

Report Number
2050012-2011-00172
Event Type
Malfunction
Date Received
January 17, 2011
Date of Event
December 19, 2010
Report Date
December 19, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PER THE CUSTOMER, QC BEEN WITHIN ACCEPTABLE RANGE. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2010 FOR THIS EVENT. THE FSE REPLACED THE CHEMISTRY CARTRIDGE (CC) SAMPLE SYRINGE BARREL, THE 3-WAY VALVE, AND FLUSHED THE SAMPLE PROBE. THE FSE THEN FOUND TUBING LEAKING BETWEEN THE SAMPLE PROBE AND THE OBSTRUCTION DETECTION TRANSDUCER. THE FSE REPLACED THE TUBING, RE-CALIBRATED THE SYSTEM, AND RAN QC. ALL RESULTS WERE ACCEPTABLE AND THE INSTRUMENT WAS RUNNING WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO OBTAINING FALSELY LOW MAGNESIUM (MG) RESULTS THAT WERE GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE CUSTOMER INDICATED THAT THE LOW MG RESULTS RANGED <0.1 MMOL/L, WHICH WERE REPEATED ON AN ALTERNATE SYSTEM THAT RECOVERED WITHIN NORMAL RANGE. PATIENT TREATMENT WAS NOT AFFECTED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1