FDA Adverse Event Malfunction Summary report: N

COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1962150 · Received January 17, 2011

Report

Report Number
6000001-2011-00314
Event Type
Malfunction
Date Received
January 17, 2011
Date of Event
December 23, 2010
Report Date
December 23, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF "AIR DETECTOR PROBLEM." (B)(4).

Additional Manufacturer Narrative · 1

EVALUATION: THE REPORTED CONDITION OF CHANNEL B AIR DETECTOR PROBLEM WAS CONFIRMED IN THE EVENT HISTORY BUT WAS NOT DUPLICATED. THE AIR IN LINE ALARMS FUNCTIONED AS DESIGNED, THEREFORE NO CORRECTION IS NEEDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. ADDITIONAL INFORMATION: THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.03.00 CATEGORIZED AS UNREMEDIATED. (B)(4).

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. UPON COMPLETION OF BAXTER?S INVESTIGATION, A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

THE FIELD SERVICE ENGINEER REPORTED ON BEHALF OF THE FACILITY A COLLEAGUE INFUSION PUMP WITH A REPORTED CONDITION OF "AIR DETECTOR PROBLEM." IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, IT IS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SOFTWARE VERSION FOR THIS DEVICE IS CURRENTLY NOT KNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1