FDA Adverse Event Malfunction Summary report: N

SUCTION TUBE MCEN770B30 BLUNT TIP

MDR report key: 19621303 · Received June 27, 2024

Report

Report Number
3003249645-2024-00037
Event Type
Malfunction
Date Received
June 27, 2024
Report Date
August 21, 2024
Manufacturer
INTEGRA MICROFRANCE S.A.S.
Product Code
JZF
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE BLUNT TIP SUCTION TUBE (MCEN770B30) WAS RETURNED FOR EVALUATION: THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES RELATED TO THE REPORTED FAILURE WERE OBSERVED. FAILURE ANALYSIS: EVALUATION OF THE SUCTION TUBE FOUND THAT THE STEM WAS BROKEN AT THE BASE OF THE HANDLE AT THE WELD. ROOT CAUSE ANALYSIS: IT WAS DETERMINED THAT AN ERROR IN THE HANDLING OF THE INSTRUMENT BY THE CUSTOMER, AN EXCESSIVE FORCE WAS APPLIED LEADING TO STEM BREAKAGE. IT IS NOTED THAT THE DEVICE IS 19 YEARS OLD.

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT THE NUMBER OF USES FOR THIS INSTRUMENT IS UNKNOWN AND COULD NOT BE ESTIMATED. IT IS NOTED THAT THIS INSTRUMENT WAS MANUFACTURED 19 YEARS AGO.

Description of Event or Problem · 0

THIS REPORT IS 10 OF 10 AND IS RELATED TO MFG REPORT NUMBERS: 3003249645-2024-00029; 3003249645-2024-00028; 3003249645-2024-00031; 3003249645-2024-00030; 3003249645-2024-00032; 3003249645-2024-00033; 3003249645-2024-00034; 3003249645-2024-00035; 3003249645-2024-00036. IT WAS REPORTED THAT ON AN UNSPECIFIED DATE, THE WELDING OF THE BLUNT TIP SUCTION TUBE (MCEN770B30) BROKE BETWEEN THE HANDLE AND STEM. ANOTHER DEVICE WAS AVAILABLE FOR USE. NO INFORMATION ON PATIENT IMPACT IS AVAILABLE AT THIS TIME.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1722945 SUCTION TUBE MCEN770B30 BLUNT TIP PFM11 JZF INTEGRA MICROFRANCE S.A.S. 04-05

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown