HARMONIC ACE 36CM W ERG HANDLE
Report
- Report Number
- 3005075853-2011-00211
- Event Type
- Malfunction
- Date Received
- January 17, 2011
- Date of Event
- October 11, 2010
- Report Date
- January 3, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH TISSUE PAD DETACHED BUT WITH EVIDENCE OF THE TISSUE PAD MATERIAL IN THE GROOVE SECTION OF THE CLAMP ARM. THE BLADE WAS RECEIVED IN GOOD CONDITION AND NOT BROKEN OFF THE DEVICE. BASED ON THE CONDITION OF THE TISSUE PAD, A PROBABLE CAUSE FOR THIS DAMAGE IS THAT THE CLAMP OF THE DEVICE MAY HAVE BEEN CLOSED AND THE INSTRUMENT ACTIVATED WITHOUT TISSUE PRESENT. CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. BOTH CONDITIONS MAY CAUSE A SYSTEM FAILURE SIGNALED BY A CONTINUOUS BEEP WHEN EITHER OF THE FOOT PEDALS IS DEPRESSED. KEEP THE CLAMP ARM OPEN WHEN BACKCUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
PER USER FACILITY MEDWATCH FORM, (B)(4), DURING A HERNIA PROCEDURE THE DEVICE TIP BROKE. NO INJURY TO PATIENT. A NEW DEVICE WAS OBTAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36CM W ERG HANDLE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |