LIGAMAX-5MM
Report
- Report Number
- 3005075853-2011-00207
- Event Type
- Malfunction
- Date Received
- January 17, 2011
- Report Date
- October 25, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE FIRED THROUGH LOCKOUT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT HAD BEEN FIRED THROUGH. PLEASE NOTE THE DEVICE CONTAINS A LOCKOUT FEATURE THAT IS DESIGNED TO INCREASE THE REQUIRED FORCE IT TAKES TO CLOSE THE TRIGGER ONCE THE LAST CLIP HAS BEEN FIRED; THIS REDUCES THE POSSIBILITY OF EMPTY JAWS BEING CLOSED ON A STRUCTURE. IN ADDITION, IF THE TRIGGER IS FORCED CLOSED ONCE THE LOCKOUT HAS BEEN ENGAGED, THE JAWS MAY REMAIN IN THE CLOSED POSITION. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TESTING COULD BE PERFORMED TO EVALUATE THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS WOULD NOT ADVANCE. DURING A PROCEDURE AT THE END OF (B)(6) (DATE UNKNOWN), THE APPLIER WAS STUCK UPON THE FIRST USE: NO CLIP DELIVERED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | G4RE6Y |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |