HARMONIC ACE 23 CM
Report
- Report Number
- 3005075853-2011-00197
- Event Type
- Malfunction
- Date Received
- January 17, 2011
- Date of Event
- December 30, 2010
- Report Date
- December 30, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED WITH THE BLADE DISCOLORED (BLUE) DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. THE BLADE WAS FOUND TO BE BROKEN AT THE DISCOLORED (BLUE). PROBABLE CAUSE OF BLADE DISCOLORATION IS APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TITANIUM TIP OF THE SCALPEL WAS BROKEN. THE BROKEN PART WAS RETRIEVED WITH THE CLIP. CHANGED TO ANOTHER DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 23 CM | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |