FDA Adverse Event Injury Summary report: N

EXCALIBUR, 5.5MM X 13CM

MDR report key: 1962005 · Received January 17, 2011

Report

Report Number
1220246-2010-00264
Event Type
Injury
Date Received
January 17, 2011
Date of Event
December 23, 2010
Report Date
December 27, 2010
Manufacturer
ARTHREX, INC.
Product Code
GFA
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

PATIENT DEMOGRAPHICS (DATE OF BIRTH, GENDER, WEIGHT) WERE REQUESTED, BUT NOT PROVIDED. NO FURTHER PATIENT INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT. DEVICE HISTORY RECORD REVEALED NOTHING RELEVANT TO THIS EVENT. THE DEVICE WAS RECEIVED AND AN EVALUATION WAS CONDUCTED. THE COMPLAINT WAS CONFIRMED. THE EVALUATION REVEALED METAL GOUGING FOUND AT THE PROXIMAL END OF THE CUTTING WINDOW ON THE INNER SHAFT. THIS IS TYPICALLY CAUSED WHEN THE USER APPLIES EXCESSIVE BENDING/LEVERAGING FORCE ON THE DEVICE DURING USE OR FROM USER MECHANICAL DAMAGE TO DEVICE SUCH AS HITTING THE DEVICE WITH ANOTHER DEVICE OR GRINDING AGAINST A HARD SURFACE. THIS IS THE FIRST COMPLAINT OF THIS TYPE FOR THIS PART/LOT NUMBER COMBINATION. THE POTENTIAL CAUSES OF THIS EVENT ARE BEING COMMUNICATED TO THE EVENT REPORTER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SHAVER BLADE WAS SPITTING METAL SHAVINGS INTO THE JOINT. SWITCHED TO ANOTHER BLADE TO COMPLETE THE CASE. METAL SHAVINGS REMAIN UNRETRIEVED IN THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXCALIBUR, 5.5MM X 13CM BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL GFA ARTHREX, INC.

Patients

Seq Age Sex Outcome Treatment
1 Other