INTEGRATED APD SET W/CASSETTE3-PRONG
Report
- Report Number
- 1423500-2011-00691
- Event Type
- Malfunction
- Date Received
- January 17, 2011
- Date of Event
- October 13, 2010
- Report Date
- December 29, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). IT IS UNKNOWN WHETHER OR NOT THERE WAS PATIENT INVOLVEMENT AT THE TIME THE ALARM OCCURRED.
(B)(4). A SYSTEM ERROR (SE) 2240 WAS FOUND DURING A REVIEW OF THE EVENT LOGS OF THE HC DEVICE. NOT ENOUGH DATA IS AVAILABLE WITHIN THE COMPLAINT TO IDENTIFY ROOT CAUSE. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. BAXTER HAS FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM AND WILL CONTINUE TO MONITOR PRODUCT LINES FOR TRENDS AND WILL TAKE CORRECTIVE/PREVENTATIVE ACTION AS APPROPRIATE. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER, THEREFORE THE DEVICE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED.
DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND A SYSTEM ERROR 2240 IN PATIENT LOGS WITH OCCURRENCE DATE OF (B)(6) 2010 DURING DRAIN 1.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3-PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |