RENAL-DISPOSABLE
Report
- Report Number
- 1423500-2011-00667
- Event Type
- Malfunction
- Date Received
- January 14, 2011
- Date of Event
- December 21, 2010
- Report Date
- December 22, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
(B)(4). THIS REPORT WAS NOT CONFIRMED DUE TO THE LACK OF A SAMPLE. BASED ON THE INFORMATION GATHERED DURING BAXTER'S INVESTIGATION, THE CAUSE OF THE ALARM WAS NOT DETERMINED. THE LOT INFORMATION WAS UNKNOWN; THEREFORE, A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THE SAMPLE IS NOT AVAILLABLE AND THE LOT NUMBER IS UNKNOWN. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THEN A FOLLOW UP MDR WILL BE SUBMITTED.
A CUSTOMER CONTACTED BAXTER'S GLOBAL TECHNICAL SERVICE CENTER TO REPORT A SYSTEM ERROR (SE) 2240 ALARM (INDICATING AIR IN THE SET) ON THE HOMECHOICE DEVICE DURING THE PREVIOUS NIGHT'S THERAPY. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND ADVISED THE CALL TO CONTACT THE NURSE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RENAL-DISPOSABLE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE |