FDA Adverse Event Malfunction Summary report: N

PRECISION XTRA

MDR report key: 1961867 · Received January 14, 2011

Report

Report Number
2954323-2011-00688
Event Type
Malfunction
Date Received
January 14, 2011
Date of Event
December 20, 2010
Report Date
January 14, 2011
Product Code
NBW
Removal / Correction Number
ADC FA1197-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30C DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE C, D OR E ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. THE PRIMARY CAUSE HAS BEEN IDENTIFIED TO BE BATCH TO BATCH GLUE VARIABILITY. THE FDA HAS BEEN INFORMED OF THE FIELD ACTION PER 21CFR806 (FIELD ACTION REFERENCE NUMBER 2954323-12/22/10-001-R). ALL AFFECTED CONSIGNEES WERE NOTIFIED BY LETTER BEGINNING DECEMBER 22, 2010.

Description of Event or Problem · 1

IN USING AN AFFECTED TEST STRIP RELATED TO AN ON-GOING FIELD ACTION FOR ABBOTT'S PRECISION FAMILY TEST STRIPS, THE CUSTOMER'S ON IN LAW REPORTED EITHER LONGER BLOOD FILL TIME OR READINGS ISSUES. SPECIFICALLY, THE CUSTOMER REPORTED THAT ON (B)(6) 2010 THEY RECEIVED READINGS OF 37MG/DL AND 47MG/DL RESPECTIVELY. CUSTOMER'S FAMILY MEMBER STATED AT THE TIME THE READINGS WERE OBTAINED THE CUSTOMER EXPERIENCED SYMPTOMS OF "SWEATS AND DISORIENTATION". NO THIRD PARTY MEDICAL INTERVENTION WAS REPORTED AND CUSTOMER DRANK JUICE TO HELP ALLEVIATE THEIR SYMPTOMS. THERE WAS NO REPORT OF SERIOUS INJURY, DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISION XTRA BLOOD GLUCOSE MONITORING SYSTEM NBW 45737

Patients

Seq Age Sex Outcome Treatment
1