FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1961621 · Received December 27, 2010

Report

Report Number
1720753-2010-04908
Event Type
Malfunction
Date Received
December 27, 2010
Date of Event
December 10, 2010
Report Date
December 27, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE COLLIMATOR WAS CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9800 SYSTEM HAD A PHYSICIST DISCREPANCY OF THE COLLIMATOR WAS TOO LARGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1