FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 1961556 · Received January 14, 2011

Report

Report Number
3005075853-2011-00185
Event Type
Malfunction
Date Received
January 14, 2011
Date of Event
December 16, 2010
Report Date
December 17, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION WAS REQUESTED. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS VISIBLE AT 12TH FIRING SEQUENCE, THUS IT OVER TRAVELED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP CHOLECYSTECTOMY PROCEDURE, THE CLIP OF THE DEVICE COULD NOT BE OPENED AFTER THREE FIRINGS. THEN THE SURGEON OPENED THE CLIP BY HAND AND TRIED OUTSIDE OF THE PATIENT, BUT IT STILL FAILED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1