FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 1961549 · Received January 10, 2011

Report

Report Number
3015876-2011-00037
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 15, 2010
Report Date
December 15, 2010
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE TO DETECT A THERAPY CABLE CONNECTION. PHYSIO DETERMINED THE CAUSE OF THE FAILURE TO BE A DISCONNECTED PCB MOUNTED JACK CONNECTOR, J23, TO THE THERAPY PCB ASSEMBLY. PHYSIO RESEATED THE CONNECTION AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT RECOGNIZE A THERAPY CABLE CONNECTION. THE DEVICE WOULD NOT BE AVAILABLE TO DELIVER DEFIBRILLATION THERAPY IN THE REPORTED CONDITION. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA